This is not the document you are looking for? Use the search form below to find more!

Report home > Science

EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay

0.00 (0 votes)
Document Description
Quinine is a very strongly fluorescing compound, especially in dilute acid solution, and thus can be detected in very trace amounts. In 0.05 M H 2 SO 4 , quinine has two analytically useful excitation wavelengths : λ ex = 250 and 350 nm. Regardless of which excitation wavelength is used, the wavelength of maximum fluorescence emission intensity, λ em or λ fl, is 450 nm. The basis for quantitation is that the intensity of fluorescence emission in very dilute solutions is directly proportional to the concentration of quinine - if the intensity of the excitation source and other experimental factors are kept constant.
File Details
Submitter
  • Name: harumi
Embed Code:

Add New Comment




Related Documents

CHEM 370 - Fluorescence Spectroscopy - Determination of Quinine

by: shaiming, 6 pages

Fluorescence is the emission of light by a molecule, which has absorbed radiant energy; the radiation is emitted at a longer wavelength (lower energy) than the incident absorbed energy. S o ...

FLUORESCENCE SPECTROSCOPY OF QUININE

by: frans, 5 pages

In this experiment, we will make use of a recording spectrofluorimeter to study the luminescence characteristics of the bicyclic alkaloid quinine, which is commonly used as a fluorescence standard. ...

Logo experiment 6

by: Sonja, 1 pages

Logo experiment 6

Luminescence Quenching of Quinine

by: gerolt, 5 pages

The purpose of this laboratory is to study the fluorescence of quinine and its quenching by halide ions. You will record the excitation and emission spectra of quinine in tonic water. Then you will ...

SEMI- QUANTITATIVE ANALYSIS OF QUININE SULFATE IN TONIC WATER BY SERS

by: niklas, 2 pages

Quinine sulfate has been examined by dispersive Raman spectroscopy in powdered and aqueous solution state as well as by SERS. The analytical parameters for the analysis of quinine sulfate by SERS are ...

TOLERANCE TO EFFECTS OF COCAINE ON BEHAVIOR UNDER A RESPONSE-INITIATED FIXED-INTERVAL SCHEDULE

by: shinta, 12 pages

Tolerance to effects of cocaine can be modulated by schedules of reinforcement. With multiple ratio schedules, research has shown an inverse relationship between ratio requirement and amount ...

QUININE DIHYDROCHLORIDE 6% Sterile Concentrate 600 mg in 10 mL

by: katja, 5 pages

PRODUCT INFORMATION. Quinine Dihydrochloride 6% Sterile Concentrate 600 mg in 10 mL

Laboratory 2: Determination of quinine in tonic water by ...

by: henriette, 6 pages

The purpose of this lab is to use the UV-Vis spectrophotometer to analyze the amount of quinine and sodium benzoate in tonic water and report how many 8-ounce bottles of tonic water a 140-pound ...

Chapter 6.2 : Covalent Bonding and Molecular Compounds

by: janet, 19 pages

Chapter 6.2Covalent bonding and molecular compoundsObjectives:Define molecule and molecular formula.Explain the relationships between potential energy, ...

Improving Yoghurt Quality of Goat Milk by Implemented Complete Feed Omega 6 on Goat Feeding Program

by: shinta, 2 pages

Goat milk is a potential food commodity that needs to be developed. It is for fulfilling the quality standard then must be implemented the safety, healthy, intact, clean and ...

Content Preview
Department of Chemistry

University of Kentucky
EXPERIMENT 6

Molecular Fluorescence Spectroscopy: Quinine Assay



UNKNOWN

Submit a clean, labeled 500-mL volumetric flask to the instructor so that your unknown
quinine solution may be issued. Your name, section number, and your locker number
should be written legibly on this flask. The flask does not need to be dry on the inside, but
needs to have been rinsed with distilled water after it has been cleaned. The flask must be
turned in at least 1 lab period before you plan to do the experiment
so that the Teaching
Assistants will have time to prepare the unknown. Each student will have his or her own
unknown to analyze even if you are working in pairs.


BACKGROUND

Quinine (C20H24N2O2, 324.43 g/mol) is an alkaloid extracted from the bark of the cinchona tree.
It has been used for many years as an antimalarial agent. Although it does not cure malaria, it is
effective in alleviating the symptoms of malarial attacks. The usual medicinal form is quinine
dihydrochloride or quinine sulfate dihydrate, (C20H24N2O2)2 H2SO4 2H2O, 782.97 g/mol.

Quinine is a very strongly fluorescing compound, especially in dilute acid solution, and thus can
be detected in very trace amounts. In 0.05 M H2SO4, quinine has two analytically useful
excitation wavelengths: λex = 250 and 350 nm. Regardless of which excitation wavelength is
used, the wavelength of maximum fluorescence emission intensity, λem or λfl, is 450 nm. The
basis for quantitation is that the intensity of fluorescence emission in very dilute solutions is
directly proportional to the concentration of quinine – if the intensity of the excitation source and
other experimental factors are kept constant. Because the absolute emission intensity can vary
considerably with small differences in experimental conditions, a calibration curve is prepared
by measuring the fluorescence-emission intensity of accurately known quinine standard
solutions.


INSTRUMENTATION

Turner Quantech Digital Filter Fluorometer, Model FM 109525

The instrument used in this experiment uses glass filters to select wavelength ranges appropriate
for quinine, rather than some type of expensive grating or prism monochromator to isolate
narrow excitation and fluorescence wavelengths. The latter type may typically have a bandpass
of 0.1 or 1 nm. A bandpass is defined as the width at half-height of the maximum intensity of
light that passes through the filter. It is also called the bandwidth.
CHE 226 – Analytical Chemistry Laboratory 34
Quinine
Fluorescence

Department of Chemistry

University of Kentucky

A monochromator-based instrument is much better for obtaining spectra, particularly those with
fine structure, but a filter-based unit will often be better for routine quantitative analysis. The
much larger bandpasses permit greater fractions of the excitation and fluorescent light to excite
the sample and to reach the detector, respectively. This greatly increases the sensitivity of the
instrument and can thus usually lower typical detection by an order of magnitude or more. Filter
fluorometers are also much less expensive.


Excitation Source: A 5-watt quartz-halogen lamp, which emits intense broadband radiation from
340 nm to 750 nm.
Excitation Wavelength Filter: A narrow-band 360-nm filter with a bandpass of 40 nm.
Emission Wavelength Filter: A sharp cut-in, long-wavelength-pass filter which transmits
essentially all light with λ > 415 nm.
Detector: Photomultiplier tube, model 931B PMT.
Detection Limit. Stated in the manufacturer’s literature as 30 ppt quinine sulfate, which is 30
ng/L or about 9 x 10-11 M.

Turn the instrument on at least 15 minutes before using to allow it to warm up and
stabilize
. The ON-OFF switch is on the back panel near the power cord. When turned on, the
instrument runs a countdown timer during which it undergoes self tests.


PREPARATION OF SOLUTIONS

Preparation of Stock Sulfuric Acid Solutions

The two stock H2SO4 solutions should have already been prepared for you by the Teaching
Assistants. However, if you use up all of the 0.05 M solution while doing your experiment,
prepare another 2-L batch for those that follow. It is very straightforward.

1 M H2SO4. Slowly and carefully add 56 mL conc. H2SO4 to about 500 mL distilled water in a
1-L beaker with stirring. This solution should have been already prepared for your use.

0.05 M H2SO4. With a graduated cylinder, add 100 mL of 1 M H2SO4 with a graduated cylinder
to about 500 mL of distilled water in the screw-capped acid reagent jug labeled for this solution.
Mix and dilute to the 2.0-L mark on the bottle and mix thoroughly. This solution should already
have been prepared for your use. If you use all of or most of what is there, prepare a new batch
for the students who follow you.

Preparation of Quinine Stock Solution, 1000 ppm

1. Carefully weigh exactly 0.1207 g of quinine sulfate dihydrate onto a folded glassine
weighing paper or into a small plastic weighing boat, and transfer this quantitatively into a
100-mL volumetric flask. A few squirts of distilled water from a wash bottle should help to
wash the solid material from the weighing boat and the neck of the flask.
CHE 226 – Analytical Chemistry Laboratory 35
Quinine
Fluorescence

Department of Chemistry

University of Kentucky

2. Pipet 5.00 mL of 1 M H2SO4 (located in hood #2) into the flask. Carefully dissolve all the
quinine in this sulfuric acid solution by swirling before diluting to volume. This is critically
important
.

3. Carefully dilute to volume with distilled water and mix thoroughly.

Preparation of Intermediate Quinine Stock Solution, 10.0 ppm

1. Pipet 5.00 mL of the 1000-ppm solution into a 500-mL volumetric flask.

2. Add 25.0 mL of 1 M H2SO4, dilute carefully to volume with distilled water, and mix
thoroughly.

Preparation of Quinine Standard Solutions

1. Using volumetric transfer pipets and/or a 10-mL graduated pipet, add 1.00, 3.00, 5.00, 7.50,
and 10.00 mL of the 10-ppm intermediate stock solution into five properly labeled 100-mL
volumetric flasks. This will result in standard solutions of 0.1, 0.3, 0.5, 0.75, and 1.0 ppm.

2. Carefully dilute to volume with 0.05 M H2SO4.

3. 0.05 M H2SO4 is used as the “blank.”

Preparation of Quinine Unknown

1. Your unknown solution is obtained from the teaching assistants in a 500-mL volumetric
flask. Add 25.0 mL of 1 M H2SO4 to the flask.

2. Dilute to volume with distilled water and mix thoroughly.


PROCEDURE

Turn on the Quantec fluorometer at least 15-20 minutes before making measurements in order to
let it warm up and stabilize.

Measurement of Emission Intensities

1. Carefully fill separate, clean, plastic fluorescence cuvettes (these have 4 clear sides) about ¾
full with the blank (0.05 M H2SO4), each of the five standards, and each unknown sample.
Do not touch the optical surfaces with your fingers. Instead, handle with KimWipes. Be
sure to wipe any smudges off the optical surfaces of the cuvettes.

2. On the main menu of the fluorometer, press ENTER.

CHE 226 – Analytical Chemistry Laboratory 36
Quinine
Fluorescence

Department of Chemistry

University of Kentucky
3. Use the right cursor (®) to select “Quinine”. Press RETURN.

4. The instrument asks if you wish to change the name. Select NO.

5. The instrument asks if filters are correct. Select YES.

6. The instrument asks if you would like to you as standard curve from memory. Select NO.

7. The instrument asks for the number of points for the calibration curve. Use the up-arrow ( )
to increase the number to 5 in order to be able to use your five standards. Press RETURN.

8. Beginning with the highest concentration standard, enter the concentration using the up-
arrow to change the values. Use (¬) to move the cursor to the appropriate position. Use the
up-arrow to set the appropriate units (ppm).

9. The instrument asks to insert the sample. Do so and press RETURN.

10. Repeat for other standards.

11. Instrument asks the operator to insert the BLANK. Do so and press ZERO.

12. The instrument gives a “Coefficient of Determination” value. If this value is less than 0.90,
you must pour out your standard solutions and repipet them from the stock quinine solution.
If this value is greater than 0.90, continue.

13. Insert the unknown and Press RETURN.

14. Readout shows READ and provides the value.

15. Repeat the entire process at least two more times to obtain a set of three “good” replicate
measurements: If the values for a set of standards and the unknown do not “drift” over time,
this would mean that the raw fluorescence data for your unknown are within a range of about
3-4% relative. If the values do appear to drift, then the ratios of the fluorescence intensities
of the standard to those of a nearby standard are within a range of about 3-4% relative.

Comparison of Calibration Procedures

This procedure will compare values that are (a) calculated by the instrument based on its internal
calibration procedure with (b) those obtained based on the calibration curve(s) that you prepare
from the raw fluorescence-emission data obtained above.

1. At main menu, select RAW FLUORESCENCE as the units of measure.

2. The instrument requests a value for the standard. Enter the value and RETURN.

CHE 226 – Analytical Chemistry Laboratory 37
Quinine
Fluorescence

Department of Chemistry

University of Kentucky
3. The instrument instructs you to insert the standard. USE the 1.0 ppm STANDARD. Do so
and press RETURN. The instrument adjusts its internal gain.

4. The instrument asks “Insert Blank Sample?” Select YES, followed by RETURN.

5. Insert the blank and the instrument will take a reading.

6. Now insert each of the standard solutions and record the emission value for each.

7. Insert the unknown, and record the emission value.

8. Repeat the entire procedure at least twice more in order to obtain at least 3 “good”
measurements for each sample.

9. Prepare calibration curves and calculate the average concentration of the unknown sample.

When done with the entire experiment, rinse out all the plastic cuvettes with distilled water,
shake off excess water, and return them to the experiment drawer so they can be reused.


DATA ANALYSIS

Prepare appropriate calibration curve(s) from your replicate sets of data. Depending on the nature
of the data and any drift in the instrument, it may be best to (a) average all the net emission
intensities for a particular solution and to prepare one calibration curve or (b) prepare several
calibration curves from the separate sets of data, obtain several values for the concentration of
quinine in your unknown, then average these values. Determine which one is better or if both
give equivalent results. Use linear least squares to fit each of the three sets of data, and the
average of all three, in your analysis.

Report the “best value” for the concentration of quinine (in ppm) in your unknown and the
estimated standard deviation of this best value.

Note: By definition, parts-per-million (ppm) is a mass-ratio unit – ganalyte/gsample x 106 or µg/g.
In dilute aqueous solutions, 1 ppm is equivalent to 1 µg/mL or 1 mg/L because the density of the
solution is 1.00 g/mL. A 1000-ppm solution of quinine sulfate (1.000 g of quinine sulfate per
liter of solution) would only be an 828.7 ppm of quinine itself. If you cannot weigh out exactly
0.1207 g of quinine sulfate, try to get it close, and simply calculate the correct values of the
concentrations of your standards.


HAZARDOUS WASTE DISPOSAL

Dispose of ALL waste quinine and sulfuric acid solutions in the proper, labeled Hazardous
Waste Container for this experiment, which is located in the hoods. 0.05 M H2SO4 has a pH

CHE 226 – Analytical Chemistry Laboratory 38
Quinine
Fluorescence

Department of Chemistry

University of Kentucky
of about 1 and thus may NOT simply be run down the drain, even if you run large amounts of
water with it. If you are unsure about the container, ASK.

TEXT REFERENCE


D. A. Skoog, D. M. West, F. J. Holler, and S. R. Crouch, Analytical Chemistry: An Introduction,
7th ed., Chapter 23, pp. 594-631.













































Revised January 24, 2005


Copyright © by The Department of Chemistry, University of Kentucky, 2005
CHE 226 – Analytical Chemistry Laboratory 39
Quinine
Fluorescence

Download
EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay

 

 

Your download will begin in a moment.
If it doesn't, click here to try again.

Share EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay to:

Insert your wordpress URL:

example:

http://myblog.wordpress.com/
or
http://myblog.com/

Share EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay as:

From:

To:

Share EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay.

Enter two words as shown below. If you cannot read the words, click the refresh icon.

loading

Share EXPERIMENT 6 Molecular Fluorescence Spectroscopy: Quinine Assay as:

Copy html code above and paste to your web page.

loading